USES锛
銆銆This equipment is widely used for drying, curing, melting and heating in experiments of industrial enterprises, scientific research institutions, and health and medicine units etc.
CHARACTERISTICS锛
1. Vertical type, cold rolling steel electrostatic spraying exterior.
2. Well coordinated bottom air blower and heater and vertical hot air circulation ensure the temperature evenness.
3. High quality mirror finished stainless steel interion; foursquare semicircle transition; flexible positioning and removable shelf; airduct lateral plate and bottom heater covering are easy to assembly and disassembly for convenient cleaning.
4. PID micro-computer intelligent temperature controller with functions of timing and over-temperature protection. Precise in temperature control and easy to read.
5.New desighed door ,and with double-galss observation window ensures effeetive heat-dissipation, adjustable airtight buckle lock, and heat resistant silicon rubber seal, which efficiently prevent heat loss.
Note: model WGL-625B features top blowing.
heat insulation well.
MAIN TECHNICAL PARAMETERS锛
Model |
voltage
锛圴锛
|
Temperature
variation
锛堚剝锛 |
Tem.fluctuation
锛堚剝锛
|
Tem.distingu-
ishability
锛堚剝锛
|
Ambient tem.
锛堚剝锛 |
Heating power
(kw)
|
WGL-30(B) |
220V/50HZ |
RT+10-300鈩 |
±1.0鈩 |
0.1鈩 |
+5-40鈩 |
0.8 |
WGL-45(B) |
1.2 |
WGL-65(B) |
1.6 |
WGL-125(B) |
2.3 |
WGL-230(B) |
3 |
WGL-625B |
380V/50HZ |
5 |
Model |
Interior
dimensions
W×D×H(mm) |
Exterior
dimensions
W×D×H(mm) |
Packaging
dimensions
W×D×H(mm) |
Load per rack |
Gross/net
weight(kg) |
Timing
range |
WGL-30(B) |
310×310×310 |
460×510×695 |
550×570×765 |
15/kg 2 for stand-ardconfiguration |
33/37 |
1-9999min
|
WGL-45(B) |
350×350×350 |
500×550×735 |
620×585×800 |
37/43 |
WGL-65(B) |
400×360×450 |
550×550×840 |
640×640×905 |
44/49 |
WGL-125(B) |
500×450×550 |
636×680×915 |
730×720×1000 |
60/66 |
WGL-230(B) |
600×500×750 |
730×670×1220 |
900×800×1350 |
25/kg 2 for standard configuration |
94/120 |
WGL-625B |
660×760×1250 |
895×925×1830 |
1000×1080×2000 |
25/kg 3 for standard configuration |
90/160 |